Clarity - Getting Started First “Analysis”
41
5.6.1 Opening a calibration file
When a calibration file was specified in the
template method used for acquiring the
chromatogram, you can open it from the
Chromatogram window.
• Use the View button in the
Chromatogram -
Results
window.
• Use the
File - Open Standard (yellow icon)
command to open an integrated chromatogram
representing a calibration standard.
• Click the Add All
(blue line) icon to
automatically transfer available data on each
peak to the calibration table.
• In the displayed calibration table, modify
compound names in the first column
.
Note: Table fields can be edited after clicking on them
with the mouse.
• Fill in the known quantity for each compound
in the Amount column
under the number of
the current calibrated level.
Note: If you have performed all steps as described, you
should see a straight-line single-point calibration
curve after clicking on the tab of any individual
compound.
• Finally, save the created calibration curves
using the File - Save command.
5.6.2 Link the calibration file with chromatogram or template
method
Comments
The calibration can be attached (applied) to a
chromatogram by selecting its file in the
Chromatogram – Results window (in Calibration
File (Peak Table) field.