Clarity Troubleshooting
118
Analysis dialog. The interrupted analysis will then be saved as a valid
chromatogram.
Disk full
If a requested operation cannot be performed because the disk is full,
you have several options:
Delete files that are no longer needed from any directories other than
Clarity.
Archive or delete unnecessary chromatograms or entire projects.
With the Archive command in the Instrument window, open the
Backup – Create Archive dialog. Either archive the older chromatograms
on another disk or erase them using the Delete Selected Files
command.
10.3 Sequence
The Run command and icon are inactive (dimmed)
A Single Analysis is running. Check the status in the
Instrument window.
Single Analysis can be terminated from the
Single Analysis dialog or Data
Acquisition window.
See the first bullet of the chapter 10.2 - Signal Displaying and
Measurement.
10.4 Processing and Displaying
Chromatograms
A saved chromatogram has been "lost".
Chromatograms are either stored in the data or calibration subdirectory
depending on the setting of the Calibration Standard parameter in the
Single Analysis dialog. When selecting a chromatogram do not forget to
display the corresponding subdirectory using the
or
icons.