Clarity Calculations and Calibration
95
whole calibration process. The amount of the calibrated compound is
entered in Amount field in the displayed dialog.
6.8 Recalibration
The Clarity station provides for facile recalibration (modification of
calibration files). The procedure is similar to that which has already
described for calibration.
What does recalibration mean?
Recalibration means repeated calibration at a given concentration level.
The purpose is either to update the calibration points or increase
precision by repeated averaging at individual calibration levels.
Fig. 29. The Calibration Window Toolbar
How is the recalibration mode set?
The mode of recalibration is set in Mode - Recalibration field in the
Calibration Options dialog. The command can be opened by invoking the
Calibration - Options command, using the
icon, or directly in the
main toolbar of the calibration window.
Fig. 30. The Calibration Options Dialog
How to set the recalibration mode?
Set the recalibration mode in the Recalibration section of the
aforementioned
Calibration Options window. The result of recalibration