Clarity Chromatogram Measurement and Evaluation
42
4.3.1.2 Continuous (Preliminary) Measurement Processing
While an analysis is running the already acquired part can be evaluated
(while the analysis continues) using the Snapshot command from the
Single Analysis or Data Acquisition windows. This command is almost
identical to the Stop command but does not stop the analysis. The
Snapshot is suitable e.g. for a preliminary evaluation of a prolonged
analyses.
The Snapshot does not increment the counter of the analyses (%n) and
a warning message regarding an eventual overwrite of the existing
chromatogram will be displayed only when the Snapshot command has
been used in the analysis for the first time. Any changes will then be
overwritten by next use of the Snapshot command or after the end of
the analysis in progress.
4.3.1.3 Control of External Devices
The Clarity station (when using A/D converter) is able to control
external devices equipped with control inputs (pumps, temperature
programmers, autosamplers, etc.). The Event Table, included in each
method, can control up to eight inputs on the internal or four inputs on
the external A/D converter. Each line of the table defines a single
condition - event - and, if the condition is satisfied, a set a specified
output to the prescribed value. Examples of events: the expiration of a
preset analysis time - Time, the decrease of a detector signal below or
increase above, a preset value measured either from zero - Input or
from the value at the start of the current analysis - Input Rel.
When contradictory requirements involving a single output occur, the
last command issued will be in control.
When using a multi-detector measurement there will be a separate
Event Table controlling the outputs of the A/D converter and
containing the corresponding channel for each detector.
Event Table creation and modification
To display the Event Table invoke the
Method Setup – Event Table dialog
command from the
Instrument window.